Starting BSS
<Caution!> PCs at the beamlines should be used for reseach purpose only. When detected any uses for the other purposes, e.g. the connection to undesirable sites, the network connection will be automatically down for 30 minutes. The beamline control system will malfunctions.


You can start the beamline control software BSS by clicking the BSS icon on the desktop.
Please note that you cannot start more than two BSS processes at the same time

First of all, BSS asks whether the sample changer SPACE will be used or not. Please select the "No" button if you mount samples by hands.

Next, the Question dialog will appear and ask if you want to initialize the related devices: the rotational axis of the attenuator, rotational and translational axes of the goniometer, beam stopper, collimator, and camera length of the current detector.

This is to warn you that if you have already mounted your crystal, initialization of the goniometer axes will move your crystal out of the stream of cold gas If no crystal is mounted, press OK for initialization to proceed. If a crystal is already mounted, press Cancel. Although measurement is possible without initialization, accuracy of the starting position of goniometer and other values is not guaranteed. Between measurements, please dismount the crystal and perform initialization (click Setting, Initialize).

Completing the initialization process, BSS shows ”End station initialization finished.” in the System Message tab of Message Console. Now you can start measurement.